PH100-Si-HA-D0

P/N 202681

Photodiode detector for laser power measurement up to 36 mW.

    • 5 mm Ø
    • 7 mm Ø
    • 10 mm Ø
    • Silicon
    • UV-Silicon
    • Germanium
    • None
    • OD1
    • OD2
    • OD0.3
    • Standard
    • INTEGRA (USB)
    • INTEGRA (RS-232)
    • INTEGRA (USB with external trigger)
    • BLU (wireless)
PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0 PH100-Si-HA-D0

Traceable calibration:

Logo nist

Available with

Logo Integra

Product family key features

Large apertures

10 mm Ø for the silicon sensors

3 versions

  • Silicon 350 - 1080 nm, up to 750 mW
  • Silicon-UV 210 - 1080 nm, up to 38 mW
  • Germanium 800 - 1650 nm, up to 500 mW

Choice of attenuators

Models with attenuators include a calibration both with and without the removable filter

High accuracy

The new PH100-SI-HA presents the lowest calibration uncertainty to date.

Precise calibration

Wavelength selection in 1 nm steps

Smart interface

Containing all the calibration data

Compatible stand

Compatible displays & PC interfaces

  • MIRO ALTITUDE MIRO ALTITUDE

    Touchscreen display device for power & energy measurement. Single-channel.

  • MAESTRO MAESTRO

    Touchscreen display device for power & energy measurement. Single-channel, multiple outputs (USB, Ethernet, analog).

  • U-LINK (USB) U-LINK (USB)

    PC interface for power & energy measurement. Single-channel, USB output.

  • U-LINK (RS-232) U-LINK (RS-232)

    PC interface for power & energy measurement. Single-channel, RS-232 output.

  • INTEGRA INTEGRA

    Embedded PC interface for power & energy measurement. Can only be purchased as a detector output option.

  • TUNER TUNER

    Display device with tuning needle for power measurement. Single-channel.

  • UNO UNO

    Economical display device for power measurement. Single-channel.

  • P-LINK (USB) P-LINK (USB)

    PC interface for power measurement. Single-channel, USB output.

  • P-LINK (RS-232) P-LINK (RS-232)

    PC interface for power measurement. Single-channel, RS-232 output.

  • M-LINK M-LINK

    PC interface for power & energy measurement. Single-channel, USB output.

  • P-LINK-4 (USB) P-LINK-4 (USB)

    PC interface for power measurement. Four channels.

Included

  • Calibration certificate

  • Personal wavelength correctionTM Certificate

  • Protective cover for optical surfaces

Measurement capabilities

  • Maximum average power1

    36 mW
  • Noise equivalent power2

    10 pW
  • Spectral range

    350 - 1080 nm
  • Typical rise time

    0.2 s
  • Power calibration uncertainty

    ±5.0 % (350 - 399 nm)
    ±2.0 % (400 - 449 nm)
    ±1.5 % (450 - 809 nm)
    ±2.0 % (810 - 899 nm)
    ±4.0 % (900 - 1009 nm)
    ±7.5 % (1010 - 1080 nm)
  • Peak sensitivity

    0.5 A/W @ 980 nm
  • Minimum repetition rate3

    155 kHz
    • 1. At 1064 nm. See curves for maximum power at other wavelengths.
    • 2. At 980 nm. Nominal value. Actual value depends on environmental electromagnetic interference and wavelength.
    • 3. See user manual for details.

Damage thresholds

  • Maximum average power density

    100 W/cm²

Physical characteristics

  • Aperture diameter

    10 mm
  • Absorber

    Si
  • Dimensions

    38.1Ø x 27.4D mm
  • Weight

    0.13 kg
  • Distance to sensor face

    13.7 mm

CA$844.00*

Measurement capability in the picowatt level

Photodiode-based detector offering a fast response for VIS-NIR wavelengths. Our silicon absorber makes it possible to measure very low powers down to the picowatt level.

Customers enjoy:

  • Wide dynamic range
  • Very low noise
  • Fast response
  • Not sensitive to the surrounding environment

NIST-TRACEABLE CALIBRATION INCLUDED

Traceable calibration provides the ability to trace the readings of laser output measurements to known international measurement standards determined by the National Institute of Standards and Technology (NIST). This ensures that the measurements taken by the instruments are reliable and accurate.

Benefits include:

  • Make valid comparisons of measurements taken by other instruments inside and outside your organization and over time
  • Comply with regulations or conform to industry standards
  • Avoid costly measurement mistakes in your operations
  • Improve customer confidence in your laser systems
  • Ensure laboratory experiment validity
  • Demonstrate measurement accuracy easily with the calibration certificate we provide
 

Optional INTEGRA (USB or RS-232) output

When you choose the INTEGRA output option you get an all-in-one measurement instrument that combines both the detector and the meter in one convenient product. Plug the detector in your PC, install our free software and start measuring.

Why choose it?

  • The most cost-efficient solution
  • Same measurement performance as our other solutions
  • No compromise on interface functions and features
  • Compact solution for OEM applications and field servicing
  • Save money on recalibration with an all-in-one instrument
 

Related videos

All videos

Compatible products you may need

*Does not include shipping fees, import fees or local representation. Request quote for full details and local sales contact.

Your item has been added to your cart!

You might need these specific products to be able to use the placeholder only.

Compatible products

No compatible product

Check if this product works for your laser

Measurement type:

1

Beam size

2

Laser type

3

Beam profile

COPYRIGHT ©2024 Gentec-eo Spektrum média