UP19K-15S-VR-D0

P/N 201147

Thermal detector for laser power measurement up to 15 W.

    • 10 mm Ø
    • 12 mm Ø
    • 16 - 19 mm Ø
    • 25 mm Ø
    • 50 - 55 mm Ø
    • Convection (up to 15 W)
    • Heatsink (up to 30 W)
    • Large Heatsink (up to 50 W)
    • Fan (up to 110 W)
    • Water (up to 200 W)
    • Convection - compact casing (up to 6 W)
    • Water – large casing
    • H - Broadband
    • W - Very high power density
    • QED - High energy density
    • F - IR filter
    • HD - Broadband, high power
    • VP - Ultrashort Pulses
    • Standard
    • INTEGRA (USB)
    • INTEGRA (RS-232)
    • INTEGRA (USB with external trigger)
    • BLU (wireless)
UP19K-15S-VR-D0

Traceable calibration:

Logo nist

Available with

Logo blu Logo Integra

Product family key features

This product has been superseded. Check this page for the new product.

Modular Concept

Increase the power capability of your detector: 2 different cooling modules

High Peak Power Volume Absorber

Perfect for high density beams

Compact Design

Only 21 mm thick (15S model)

Energy Mode

Measure single shot energy up to 40 J

Smart Interface

Containing all the calibration data

Compatible stand

Compatible displays & PC interfaces

  • MIRO ALTITUDE MIRO ALTITUDE

    Touchscreen display device for power & energy measurement. Single-channel.

  • MAESTRO MAESTRO

    Touchscreen display device for power & energy measurement. Single-channel, multiple outputs (USB, Ethernet, analog).

  • U-LINK (USB) U-LINK (USB)

    PC interface for power & energy measurement. Single-channel, USB output.

  • U-LINK (RS-232) U-LINK (RS-232)

    PC interface for power & energy measurement. Single-channel, RS-232 output.

  • INTEGRA INTEGRA

    Embedded PC interface for power & energy measurement. Can only be purchased as a detector output option.

  • BLU BLU

    Embedded wireless interface for power measurement. Can only be purchased as a detector output option.

  • TUNER TUNER

    Display device with tuning needle for power measurement. Single-channel.

  • UNO UNO

    Economical display device for power measurement. Single-channel.

  • P-LINK (USB) P-LINK (USB)

    PC interface for power measurement. Single-channel, USB output.

  • P-LINK (RS-232) P-LINK (RS-232)

    PC interface for power measurement. Single-channel, RS-232 output.

  • S-LINK-1 (Ethernet) S-LINK-1 (Ethernet)

    PC interface for power & energy measurement. Single-channel, USB & Ethernet output.

  • S-LINK-2 (Ethernet) S-LINK-2 (Ethernet)

    PC interface for power & energy measurement. Dual-channel, USB & Ethernet output.

  • M-LINK M-LINK

    PC interface for power & energy measurement. Single-channel, USB output.

  • S-LINK-1 S-LINK-1

    PC interface for power & energy measurement. Single-channel, USB output.

  • S-LINK-2 S-LINK-2

    PC interface for power & energy measurement. Dual-channel, USB output.

  • P-LINK-4 (USB) P-LINK-4 (USB)

    PC interface for power measurement. Four channels.

Included

  • Calibration certificate

  • Personal wavelength correctionTM Certificate

  • Protective cover for optical surfaces

Measurement capabilities

  • Maximum average power (continuous)

    15 W
  • Maximum average power (1 minute)

    20 W
  • Noise equivalent power1

    2 mW
  • Spectral range2

    0.3 - 2.5 μm
  • Typical rise time3

    2.5 s
  • Power calibration uncertainty4

    ±2.5 %
  • Repeatability

    ±0.5 %
    • 1. Nominal value, actual value depends on electrical noise in the measurement system.
    • 2. This spectral range refers to the calibration traceability.
    • 3. With anticipation.
    • 4. Including linearity with power.

Measurement capabilities (Energy mode)

  • Maximum measurable energy1

    40 J
  • Noise equivalent energy2

    0.02 J
  • Minimum repetition period

    4.5 s
  • Maximum pulse width

    90 ms
  • Energy calibration uncertainty3

    ±5 %
    • 1. For 360 μs pulses. Higher pulse energy possible for long pulses (ms), less for short pulses (ns).
    • 2. Nominal value, actual value depends on electrical noise in the measurement system.
    • 3. When single-shot energy calibration is purchased

Damage thresholds

  • Maximum average power density1

    700 W/cm²
  • Maximum energy density2

    6 J/cm²
    • 1. At 1064 nm, 10 W CW. May vary with wavelength and average power.
    • 2. At 1064 nm, 7 ns, 10 Hz. May vary with wavelength and pulse width.

Physical characteristics

  • Cooling

    Convection
  • Aperture diameter

    18 mm
  • Absorber

    VR
  • Dimensions

    50H x 50W x 20.6D mm
  • Weight

    0.16 kg

Faster measurement with XNR AnticipationTM

Measure laser power up to 10x faster (model-dependant) with XNR AnticipationTM. Our proprietary algorithm looks at the rate at which the signal increases or decreases and combines this with calibration data to predict the result. The measurement is therefore calculated and displayed faster than the natural response of the detector.

Why does it matter?

  • Save several seconds to tens of seconds for each measurement
  • Shorter operation interruption
  • Faster laser output diagnostics
  • Increased productivity
  • No compromise on measurement accuracy

Related videos

All videos

Compatible products you may need

Your item has been added to your cart!

You might need these specific products to be able to use the placeholder only.

Compatible products

No compatible product

Check if this product works for your laser

Measurement type:

1

Beam size

2

Laser type

3

Beam profile

COPYRIGHT ©2024 Gentec-eo Spektrum média